AVS2013 Session MS-TuP: Aspects of Manufacturing Science and Technology Poster Session

Tuesday, October 29, 2013 6:00 PM in Room Hall B

Tuesday Evening

Time Period TuP Sessions | Topic MS Sessions | Time Periods | Topics | AVS2013 Schedule

MS-TuP-2 Investigation on Environment Concerns of Scanning Electron Microscopy (SEM) for Nanomanufacturing Application
Fan-Chun Hsieh, Ping-Hung Lin, Chien-Yao Huang, Nien-Nan Chu, Jiann-Shiun Kao (National Applied Research Laboratories, Tawain, Republic of China)
Scanning electron microscopy (SEM) is used for surface morphology measurement of thin films in nanomanufacturing such as: atomic layer deposition (ALD), nanosphere lithography, and nanoimprint et al. For those nanomanufacturing applications, the surface morphology of thin films could affect the desired properties. As a result, the quality of the image of surface morphology must be evaluated. In this study, the effects of microwave intensity and vibration frequency are investigated. For accelerating voltage of 15kV at magnification of 150,000, the microwave intensity of alternative current (ac) of electric equipment must below 90nT. However, at 15kV accelerating voltage and 50,000 magnifications, the microwave intensity of ac must below 70nT. Besides, the allowable amplitude increases with vibration frequency between 2Hz and 5Hz. The maximum amplitude was 6.0μm (peak to peak) at 5Hz of vibration frequency. The proposed investigation can provide a suggestion to ensure the accuracy and stability of measurement for SEM.